kemal 5164124438 v2 methylation QC validated on minfiDataEPIC and against sesame (K-026)
beta r 0.99998 vs sesame 1.28.1 (max |dbeta| 0.011), detection calls 99.1% identical,
noob within 0.5% median; replicate r 0.997; all 21 control metrics pass. Spec deviations
recorded (5-field control lines, pOOBAH formula, minfi sheet keys). User guide section.
2026-09-29 00:41:54 +01:00

54 lines
3.2 KiB
Markdown
Raw Blame History

This file contains ambiguous Unicode characters
This file contains Unicode characters that might be confused with other characters. If you think that this is intentional, you can safely ignore this warning. Use the Escape button to reveal them.
# v2 module validation — methylation control QC (`simurg methyl qc`)
Date: 2026-09-29. Development on the Raspberry Pi; oracle runs on AWS (c7g.xlarge, five short runs, ≈ $0.025).
Spec: MET-001 … MET-018 (and QC-004/QC-006, QC-062).
## Data
* **minfiDataEPIC** (Bioconductor, Artistic-2.0): 3 EPIC v1 arrays of NA12878 (IDATs + minfi-style sample sheet).
* **Infinium MethylationEPIC v1.0 B5 manifest (CSV)**: 865 918 probes (142 158 Infinium I: 92 212 red, 49 946 green),
635 control beads. Read in 8.6 s on the Pi.
## Delivered
`simurg methyl qc`, `simurg methyl convert-manifest`, `simurg qc controls --array-type methylation`, methylation
datasets in `simurg qc report`; modules `formats/methyl_manifest.py`, `signal/methyl.py`, `pipeline/methyl.py`;
the 21 methylation metrics in `qc/controls.py`. User guide: "Methylation arrays".
## Results on minfiDataEPIC
| Check | Result |
|---|---|
| Run time, 3 samples, 3 workers (Pi) | 25 s (2.9 s signal processing per sample) |
| 21 control metrics | all evaluable, all pass the `methylation_epic` profile (e.g. staining green 57, BS conversion I green 13, BS conversion II 5.3, specificity II 24) |
| Probes passing detection (p < 0.05, analytical CG + SNP probes) | 90.9 %, 91.1 %, 91.3 % |
| β of the three NA12878 replicates | Pearson r 0.997 – 0.9974 between replicates; bimodal distribution |
| Sheet `Sample_Group` → PCA colours; outputs of MET-013 | present; folder loads in `qc report` |
## Against the sesame oracle (sesame 1.28.1, same probe set via `probes.tsv`/`controls.tsv`)
`scripts/v2/sesame_oracle.R` + `scripts/v2/methyl_oracle_compare.py`, bead-count masking off on both sides (sesame's
`min_beads` masks every Infinium II probe because their M address is NA), 865 918 probes per sample:
| Quantity | Agreement (3 samples) |
|---|---|
| β values | Pearson r 0.99998; \|Δβ\| median 0.0009, 99th percentile 0.008, max 0.011 |
| noob-corrected M/U signals | relative difference median 0.5 %, max 5 % |
| Detection calls at p < 0.05 | 99.0 – 99.1 % identical; passing fraction 91.0 – 91.4 % (simurg) vs 92.0 – 92.3 % (sesame) |
| Detection p-values | not identical (median \|Δp\| 0.0008) |
The remaining difference is systematic (same size in all three samples) and is not the negative controls: sesame did
not attach the supplied control table (`negControls` failed, so its background is out-of-band Infinium I signal only),
and simurg without negative controls is not closer (noob difference 0.56 % instead of 0.53 %). The exact cause is
open (K-026). For QC use — pass/fail per sample, detection rate, β for a PCA — the agreement is sufficient; simurg's
methylation numerics are documented as its own [OURS] implementation of the published methods (MET-010), not as a
bit-exact port.
## Spec deviations (K-026)
* MET-002: the vendor EPIC manifest has a 5th field (`AVG`) on some `NEGATIVE` control lines; ≥ 4 fields are
accepted and the first four used. Fewer than 4 still gives `E_METH_CONTROLS_FORMAT`.
* MET-010: the detection p-value follows sesame's pOOBAH (p = 1 − max(F(M), F(U)), each in its own channel) instead
of the ambiguous "1 − ECDF(M + U)", so an open-source oracle exists.
* MET-004: `Sentrix_ID`/`Sentrix_Position` (minfi/sesame sheets) accepted as a third key-column pair.